The Effect of Total Fields' Area and Dose Distribution in Step and Shoot IMRT on Gamma Passing Rate Using OCTAVIUS 4D-1500 Detector Phantom

المؤلفون : Abdulrahman Mohammed Abdulbaqi 1 Siham Sabah Abdullah 2 Haydar Hamza Alabedi 3 nabaa alazawy 4 Mustafa Jabbar Al-Musawi 5 Ahmed faris Heydar 5

الملخص

Introduction: Quality assurance is necessary for every IMRT plan.‎Octavius 4D-1500 detector phantom is ‎one of the new phantoms for ‎determining the treatment plan quality‎.‎ This study aimed to examine the IMRT plans using the ‎Octavius 4D-1500‎‏ ‏‎to determine if ‎it is a reliable, ‎dependable, and durable. Material and Methods: IMRT QA conducted for 30 cases: HN and pelvis. The Monaco TPS used for treatment planning. The treatment plans ‎were then applied ‎to the Octavius 4D-1500 phantom (virtually and actually), ‎‎the γ-index was calculated in VeriSoft program to evaluate the IMRT plans. Results: Significant differences were observed between the ‎‎‎measured and calculated dose distributions for HN and pelvic plans, while the treatment ‎sites did not affect the GP rate. The results of the global Gp ‎ were higher than the local GP, ‎regardless of the study criteria. The ‎HN plans ‎showed a more significant difference ‎than the pelvic plans. The HN plans, ‎a ‎strong significant correlation was found between the total ‎fields’area and %GP in both global ‎‎and local analyses, while in the pelvic plans, there was only a significant association with the ‎local %GP. Conclusion: The measured dose distributions significantly different from calculated distributions. The relationship between the ‎‎fields’ ‏‎ area ‎and ‎‎‎%GP was inverse. In the HN plans, a ‎significant correlation found between the total ‎‎fields’ ‎area ‎and %GP in both global and ‎local, while only local %GP in the pelvic plans was significant correlation. Overall, the Octavius 4D-1500 detector phantom ‎might be applicable for ‎assessing the QA of IMRT plans.

تاريخ النشر 2021-07-04
الحالة معتمد